Multimodal Comparison of Left Ventricular Function and Volume Assessment Methods in Patients With Coronary Artery Disease: Echocardiography, Nuclear Myocardial Perfusion Imaging, and Cardiac Magnetic Resonance
iRADIOLOGY 2026, 4(5): 401-411
Published: 12 October 2026
Abstract PDF (2.7 MB) Collect
Downloads:0
Background

To evaluate the correlations and agreement between echocardiography, nuclear myocardial perfusion imaging (MPI), and cardiac magnetic resonance (CMR) for assessing left ventricular (LV) function and volumes in patients with coronary artery disease (CAD).

Methods

Forty‐eight patients with suspected or confirmed CAD who underwent echocardiography, nuclear MPI, and CMR within 1 week were included. The LV ejection fractions (EF), end‐diastolic volume (EDV), and end‐systolic volume (ESV) obtained using each modality were compared using correlational and Bland–Altman analyses, with CMR serving as the reference standard. Additionally, 13N‐NH3 and 99mTc‐sestamibi (99mTc‐MIBI) MPI subgroup analyses were conducted using CMR as the reference.

Results

Nuclear MPI showed better correlation and agreement with CMR than echocardiography for EF values (r = 0.93 vs. 0.88; bias: −2.8% vs. 4.4%). For LV volumes, nuclear MPI and CMR showed excellent correlations (EDV/ESV: r = 0.94) with better ESV agreement (bias: −10.1 mL) than EDV (−25.4 mL). In subgroup analyses, 13N‐NH3 PET MPI maintained excellent correlations with CMR (EF: r = 0.95; EDV/ESV: r = 0.97), with minimal EF bias (−0.7%). 99mTc‐MIBI SPECT MPI showed a greater EF bias (−7.8%) despite a good correlation (r = 0.95) and wider limits of agreement for LV volumes (EDV: −92.3 to 47.4 mL; ESV: −66.0 to 61.7 mL).

Conclusions

13N‐NH3 PET MPI demonstrates excellent correlation and agreement with CMR, providing a reliable alternative for assessing LV function and volumes in patients with CAD. Contrarily, echocardiography systematically overestimates EFs. These findings provide practical insights into discrepancies among imaging modalities and may support clinical decision‐making in CAD management.