Open Access Clinical Research Issue
Comparison of three different instruments for vault measurements after implantable collamer lens implantation
International Journal of Ophthalmology 2026, 19(3): 475-482
Published: 18 March 2026
Abstract PDF (1.2 MB) Collect
Downloads:36
AIM

To evaluate the differences and consistency of vault measurements obtained by Scheimpflug tomography (Pentacam), anterior segment optical coherence tomography (AS-OCT, CASIA Ⅱ), and ultrasound biomicroscopy (UBM) following implantable collamer lens (ICL) V4c implantation.

METHODS

Vault measurements were acquired using three modalities: Pentacam, CASIA Ⅱ AS-OCT, and UBM. Repeated-measures analysis of variance was used to compare the vault values obtained by the three devices. The correlation and consistency of measurements among the three instruments were assessed using the Pearson correlation coefficient, intraclass correlation coefficient (ICC), and Bland-Altman plots.

RESULTS

This retrospective study enrolled 210 myopic eyes of 210 patients (158 women and 52 men) who underwent ICL implantation: 108 eyes had a myopic ICL V4c implanted, and 102 eyes had a toric ICL V4c implanted. The mean vault values measured by Pentacam, CASIA Ⅱ, and UBM were 452.64±204.20 μm, 538.57±203.54 μm, and 560.95±227.54 μm, respectively, with statistically significant differences among the three groups (P<0.05). Pearson correlation analysis showed strong positive correlations between vault values measured by different instruments (all P<0.001). ICC results indicated good consistency among the three measurement modalities (all P<0.001). Stratified analysis revealed that when the vault value was ≤250 μm, the correlation and consistency of measurements across the three instruments were lower than those in the medium and high vault subgroups.

CONCLUSION

Vault values measured by Pentacam are lower than those obtained by CASIA Ⅱ and UBM, with UBM yielding the highest mean vault values. Measurements from the three instruments are not interchangeable but can serve as mutual references due to their significant correlation and good overall consistency. Pentacam and CASIA Ⅱ demonstrate the highest consistency in vault measurement. Notably, when the vault value is ≤250 μm, the consistency between Pentacam and the other two instruments decreases significantly.