Determining Body Surface Area Using Long‐Axial Field‐of‐View Total‐Body PET: A Novel Method Without Additional Radiation Exposure
iRADIOLOGY 2026, 4(5): 393-400
Published: 24 October 2026
Abstract PDF (1.7 MB) Collect
Downloads:0
Background

Body surface area (BSA) is used to normalize physiological measures and guide chemotherapy dosing and is traditionally estimated using height‐ and weight‐based formulas. Although whole‐body computed tomography has been investigated for direct BSA estimation, BSA determination from total‐body positron emission tomography (PET) has not been reported. We developed a method to determine BSA and morphological parameters, including total body volume, from routine long‐axial field‐of‐view (LAFOV) total‐body PET (TBPET) data without additional imaging or radiation exposure.

Methods

Software was developed in C# using the open‐source Insight Segmentation and Registration Toolkit and Visualization Toolkit, with SimpleITK and ActiViz wrappers (https://simpleitk.org/, https://www.kitware.eu/activiz/). SimpleITK was used for DICOM import and body‐mask segmentation, vtkMarchingCubes for surface reconstruction, and vtkMassProperties for BSA calculation. Technical validation was performed using a routine quality‐control phantom. Clinical data from 20 patients were acquired using a uEXPLORER LAFOV TBPET scanner.

Results

PET‐derived BSA differed from the physically measured phantom surface area by 0.19%. In patients, PET‐derived BSA was strongly correlated with BSA estimated using the DuBois, Mosteller, and Haycock formulas (r = 0.962, 0.973, and 0.974, respectively). Maximum percentage differences were 6.58%, −5.84%, and −8.34%, respectively.

Conclusions

BSA can be determined from routine LAFOV TBPET data without additional radiation exposure, with potential applicability to multibed whole‐body PET. PET‐derived BSA correlated strongly with conventional estimates, although individual differences exceeded 8%. Larger studies using appropriate reference methods are needed to establish accuracy and clinical utility.

Open Access Review Article Issue
Research progress on radioactive quantum dots for biomedical applications
Nano Research 2026, 19(10): 94908298
Published: 01 September 2026
Abstract PDF (8.7 MB) Collect
Downloads:215

Quantum dots (QDs) have been widely applied in bioimaging and oncological therapeutics owing to their unique photophysical and photochemical properties. However, conventional QDs typically require excitation from an external light source, which limits their utility in the in vivo imaging and treatment of deep-seated tissues and metastatic tumours. Conversely, radionuclides can achieve deep tissue imaging and treatment without external light source, yet they often lack multifunctional applications. Radioactive quantum dots (RQDs) emerge as an integrative solution that overcomes the limitations of both constituent materials. By harnessing the decay energy of radionuclides for self-excitation, RQDs overcome the penetration barrier of conventional QDs. Simultaneously, they incorporate the rich photophysics of QDs to achieve high-resolution multimodal imaging and combination therapy, thereby circumventing the lack of multifunctionality in radionuclides. This review mainly summarises the properties of RQDs, their synthesis strategies and their applications in bioimaging and cancer therapeutics. Finally, we discuss the future potential of RQDs in the integration of tumour diagnosis and treatment, aiming to provide guidance for researchers in medicine and engineering.

Total 2