Clinical Research | Open Access

Agreement between swept-source OCT-based and Scheimpflug-based optical biometers in myopic children

Ying-Yan Qin, Zhao-Tian Zhang, Shu-Wen Xing, Shi-Ya Lin, Bai-Zhou Chen, Liang-Ping Liu( ), Yao Ni( )
State Key Laboratory of Ophthalmology, Zhongshan Ophthalmic Center, Sun Yat-sen University, Guangdong Provincial Key Laboratory of Ophthalmology and Visual Science, Guangdong Provincial Clinical Research Center for Ocular Diseases, Guangzhou 510060, Guangdong Province, China

Co-first Authors: Ying-Yan Qin and Zhao-Tian Zhang

Show Author Information

Abstract

AIM

To evaluate the agreement of axial length (AL), anterior chamber parameters, and total cornea power obtained by swept-source optical coherence tomography (SS-OCT)-based and Scheimpflug-based optical biometers in myopic children.

METHODS

AL, steep keratometry (K), flat K, posterior corneal keratometry (PK), total keratometry (TK), anterior chamber depth (ACD), horizontal corneal diameter (CD), and central corneal thickness (CCT) were obtained using IOL Master 700 and Pentacam AXL. The agreement between the devices was evaluated using intraclass correlation coefficients (ICC), Bland-Altman plots, and astigmatism vector analysis.

RESULTS

Totally 175 myopic children (48.5% male) with a mean age of 10.29±2.14y were enrolled. The ICC and Bland-Altman plots indicated a satisfactory agreement for AL, ACD, and CCT. The mean difference in CD of -0.31±0.30 mm was considered clinically significant (>0.2 mm). Additionally, measurements of K and TK obtained from the IOL Master 700 showed good agreement. Nevertheless, there were clinically significant differences observed in PK, simulated keratometry (simK), total cornea power, and astigmatism (at least 10% of the cases with a difference of >10 degrees in meridian) between the two devices.

CONCLUSION

The study findings demonstrate a significant difference in K, PK, astigmatism, and CD, indicating that the two optical biometers cannot be considered interchangeable. Therefore, it is recommended to utilize one kind device for follow-up examinations in myopic children.

References

【1】
【1】
 
 
International Journal of Ophthalmology
Pages 339-345

{{item.num}}

Comments on this article

Go to comment

< Back to all reports

Review Status: {{reviewData.commendedNum}} Commended , {{reviewData.revisionRequiredNum}} Revision Required , {{reviewData.notCommendedNum}} Not Commended Under Peer Review

Review Comment

Close
Close
Cite this article:
Qin Y-Y, Zhang Z-T, Xing S-W, et al. Agreement between swept-source OCT-based and Scheimpflug-based optical biometers in myopic children. International Journal of Ophthalmology, 2026, 19(2): 339-345. https://doi.org/10.18240/ijo.2026.02.17

761

Views

37

Downloads

0

Crossref

0

Web of Science

0

Scopus

0

CSCD

Received: 29 March 2025
Accepted: 08 August 2025
Published: 18 February 2026
© 2026 International Journal of Ophthalmology Press

This is an open access article under the CC BY-NC-ND license (http://creativecommons.org/licenses/by-nc-nd/4.0/).