Reliable Remote-Monitoring Electrochemical Potentiostat for Glucose Measurements

Yang JINHong WANGZhengliang LVShiyuan YANG( )Haoyuan CAIJunfeng JIANG
Department of Automation, Tsinghua University, Beijing 100084, China
Siemens Ltd. Corporate Technology, Beijing 100102, China
Show Author Information

Abstract

Electrochemical methods have been widely used in the chemical and pharmaceutical industries, which require accurate concentration measurements, chemical reaction detections and analyses. The electrochemical potentiostat, the core element in electrochemical instruments, have been discussed as a hot topic addressing the difficulty of applying high-precision constant voltage and picoampere current measurements. Meanwhile, reliable potentiostats are in demand for complicated industrial environments with noises as well as requirements of remotemonitors. This paper describes a potentiostat for industrial glucose measurement that is not only accurate but also fault tolerant to guarantee high reliability in industrial environments. The instrument uses standard industrial communication protocols, profibus, and a 4-20 mA current loop, for remote control and monitoring. Experimental results show that this design has 0.01% accuracy with 1 mV resolution for voltage applications and 0.01% accuracy with 1 pA resolution for current measurements. The design is also shown to be highly reliable in noisy environments.

References

【1】
【1】
 
 
Tsinghua Science and Technology
Pages 593-600

{{item.num}}

Comments on this article

Go to comment

< Back to all reports

Review Status: {{reviewData.commendedNum}} Commended , {{reviewData.revisionRequiredNum}} Revision Required , {{reviewData.notCommendedNum}} Not Commended Under Peer Review

Review Comment

Close
Close
Cite this article:
JIN Y, WANG H, LV Z, et al. Reliable Remote-Monitoring Electrochemical Potentiostat for Glucose Measurements. Tsinghua Science and Technology, 2009, 14(5): 593-600. https://doi.org/10.1016/S1007-0214(09)70122-1

71

Views

0

Downloads

1

Crossref

N/A

Web of Science

3

Scopus

19

CSCD

Received: 25 July 2008
Revised: 20 May 2009
Published: 01 June 2009
© Tsinghua University Press 2009