Original Article | Open Access

Determining Body Surface Area Using Long‐Axial Field‐of‐View Total‐Body PET: A Novel Method Without Additional Radiation Exposure

Chun‐Lei Han1 , Wen‐Shan Huang2, Jie‐Qun Yu2, Na‐Wa Xie2, Jie‐Hua Xu2 ( )
Turku PET Centre, Turku University Hospital, University of Turku, Turku, Finland
Zhuhai People's Hospital (The Affiliated Hospital of Beijing Institute of Technology, Zhuhai Clinical Medical College of Jinan University), Zhuhai, China

Chun‐Lei Han and Wen‐Shan Huang contributed equally to this study.

Show Author Information

Abstract

Background

Body surface area (BSA) is used to normalize physiological measures and guide chemotherapy dosing and is traditionally estimated using height‐ and weight‐based formulas. Although whole‐body computed tomography has been investigated for direct BSA estimation, BSA determination from total‐body positron emission tomography (PET) has not been reported. We developed a method to determine BSA and morphological parameters, including total body volume, from routine long‐axial field‐of‐view (LAFOV) total‐body PET (TBPET) data without additional imaging or radiation exposure.

Methods

Software was developed in C# using the open‐source Insight Segmentation and Registration Toolkit and Visualization Toolkit, with SimpleITK and ActiViz wrappers (https://simpleitk.org/, https://www.kitware.eu/activiz/). SimpleITK was used for DICOM import and body‐mask segmentation, vtkMarchingCubes for surface reconstruction, and vtkMassProperties for BSA calculation. Technical validation was performed using a routine quality‐control phantom. Clinical data from 20 patients were acquired using a uEXPLORER LAFOV TBPET scanner.

Results

PET‐derived BSA differed from the physically measured phantom surface area by 0.19%. In patients, PET‐derived BSA was strongly correlated with BSA estimated using the DuBois, Mosteller, and Haycock formulas (r = 0.962, 0.973, and 0.974, respectively). Maximum percentage differences were 6.58%, −5.84%, and −8.34%, respectively.

Conclusions

BSA can be determined from routine LAFOV TBPET data without additional radiation exposure, with potential applicability to multibed whole‐body PET. PET‐derived BSA correlated strongly with conventional estimates, although individual differences exceeded 8%. Larger studies using appropriate reference methods are needed to establish accuracy and clinical utility.

Graphical Abstract

References

【1】
【1】
 
 
iRADIOLOGY
Pages 393-400

{{item.num}}

Comments on this article

Go to comment

< Back to all reports

Review Status: {{reviewData.commendedNum}} Commended , {{reviewData.revisionRequiredNum}} Revision Required , {{reviewData.notCommendedNum}} Not Commended Under Peer Review

Review Comment

Close
Close
Cite this article:
Han C, Huang W, Yu J, et al. Determining Body Surface Area Using Long‐Axial Field‐of‐View Total‐Body PET: A Novel Method Without Additional Radiation Exposure. iRADIOLOGY, 2026, 4(5): 393-400. https://doi.org/10.1002/ird3.70105

3

Views

0

Downloads

0

Crossref

0

Web of Science

0

Scopus

Received: 31 March 2026
Revised: 30 May 2026
Accepted: 28 July 2026
Published: 24 October 2026
© 2026 The Author(s). Tsinghua University Press.

This is an open access article under the terms of the Creative Commons Attribution‐NonCommercial‐NoDerivs License, which permits use and distribution in any medium, provided the original work is properly cited, the use is non‐commercial and no modifications or adaptations are made.