To solve the problems of small measurement range and poor accuracy of the spectral confocal non-contact probes, a four-piecelens group with a large measurement range is designed. Furthermore, a peak wavelength extraction algorithm based on the Adam algorithm and a piecewise fitting curve calibration method are also proposed. The lens group has a displacement measurement range of 4500 μm in 470 nm-730 nm band. After the probe system is built using the lens group, it is experimentally verified that the displacement average measurement error using the Adam algorithm is 0.3 μm, which is 68.3% less than that of the centroid algorithm, 26.0% less than that of the traditional Gaussian fitting algorithm, and 24.3% less than that of the parabolic fitting algorithm, and at the same time, the displacement average measurement error using the piecewise fitting method is 92.4% less than that using the conventional polynomial fitting method. The experimental results show that the proposed probe system can achieve high-precision non-contact displacement measurement over a large range, which provides a theoretical and experimental basis for further research on high-precision displacement sensors based on spectral confocal technology.
The articles published in this open access journal are distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits use, distribution and reproduction in any medium, provided the original work is properly cited.