Magnetorheological fluid based force feedback performance for main manipulator of invasive surgical robot

Kun TANHu SHI( )Yinyun YUEXuesong MEI
School of Mechanical Engineering, Xi'an Jiaotong University, Xi'an 710049, China

Peer review under responsibility of Editorial Committee of JAMST

Show Author Information

Abstract

In view of the lack of effective feedback function for the minimally invasive surgical robot at present, the design of force feedback device and verification of force feedback effect of the master manipulator are carried out, so that the operator of the master manipulator can obtain force feedback and improve the quality of surgery. A force feedback method based on the variable shear stress characteristics of magnetorheological fluid is proposed. The telecentric mechanism of the main manipulator end and the force feedback device are designed. And the relationship model of the output torque of the force feedback device, magnetic field intensity and angular velocity is established. The output characteristics of the device is analyzed by the experiment, and the linear relationship model between input current and output force of the force feedback device is obtained. The working performance of the force feedback device is tested by simulating the force conditions of surgical instruments in different cases of prostatectomy. The experimental results show that the delay of the force feedback device is 0.1-0.2 s, and the output performance of the force feedback device is related to the type and value of the applied force signal. When the input force signals are step signal, pulse signal and small range fluctuation signal, the average error of force feedback decreases with the increase of applied force. Analyzed from the experimental results, the overall force feedback effect of the force feedback device is well, which proves its feasibility and effectiveness in assisting doctors to carry out minimally invasive surgery.

References

【1】
【1】
 
 
Journal of Advanced Manufacturing Science and Technology

{{item.num}}

Comments on this article

Go to comment

< Back to all reports

Review Status: {{reviewData.commendedNum}} Commended , {{reviewData.revisionRequiredNum}} Revision Required , {{reviewData.notCommendedNum}} Not Commended Under Peer Review

Review Comment

Close
Close
Cite this article:
TAN K, SHI H, YUE Y, et al. Magnetorheological fluid based force feedback performance for main manipulator of invasive surgical robot. Journal of Advanced Manufacturing Science and Technology, 2022, 2(2): 2022007. https://doi.org/10.51393/j.jamst.2022007

881

Views

22

Downloads

1

Crossref

4

Scopus

Received: 05 January 2022
Revised: 12 February 2022
Accepted: 05 March 2022
Published: 15 April 2022
© 2022 JAMST All rights reserved.

This is an Open Access article distributed under the terms of the Creative Commons Attribution License http://creativecommons.org/licenses/by/4.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.