Temperature Field Analysis and Simulation and Optimization of Flow Field in High-low Temperature Test Chamber

Xiaoteng Zheng1Jinping Liu1,2( )
College of Electric Power, South China University of Technology, Guangzhou, 510640, China
Guangdong Province Key Laboratory of Efficient and Clean Energy Utilization, Guangzhou, 510640, China
Show Author Information

Abstract

In this study, a 294 L high-low temperature test chamber with two power batteries in the test area is investigated. The temperature rise and heat dissipation of the batteries in the chamber at a constant temperature of 45 ℃ are simulated through CFD, and the flow field of the high-low temperature test chamber is analyzed and optimized. The results show that for the air distribution pattern on the same side as the upper air supply and lower air return, the perforated plate in the upper part of the test area is helpful in improving the uniformity of the flow field. A vortex phenomenon occurs between the two power batteries. Increasing the distance between the shelf and bottom of the test chamber and appropriately increasing the return air grille area can reduce or even eliminate this phenomenon and thereby improve the flow field uniformity. When the distance between the shelf and bottom of the inner chamber increases to 0.2 m and the height of the return air grille increases to 0.15 m, the overall inhomogeneity of the flow field in the test chamber decreases from 1.73 to 0.4, and the vortex phenomenon between the batteries disappears.

CLC number: TB61+1; TB657.3 Document code: A Article ID: 0253-4339(2024)02-0119-08

References

【1】
【1】
 
 
Journal of Refrigeration
Pages 119-126

{{item.num}}

Comments on this article

Go to comment

< Back to all reports

Review Status: {{reviewData.commendedNum}} Commended , {{reviewData.revisionRequiredNum}} Revision Required , {{reviewData.notCommendedNum}} Not Commended Under Peer Review

Review Comment

Close
Close
Cite this article:
Zheng X, Liu J. Temperature Field Analysis and Simulation and Optimization of Flow Field in High-low Temperature Test Chamber. Journal of Refrigeration, 2024, 45(2): 119-126. https://doi.org/10.3969/j.issn.0253-4339.2024.02.119

1

Views

0

Downloads

0

Crossref

0

CSCD

Received: 04 February 2023
Revised: 22 June 2023
Accepted: 29 June 2023
Published: 16 April 2024
© 2024 The Editorial Office of Journal of Refrigeration

This is an open access article under the terms of the Creative Commons Attribution 4.0 International License (CC BY 4.0, http://creativecommons.org/licenses/by/4.0/).