Research paper | Open Access

Numerical analysis of the effect of middle-ear effusion on the sound transmission and energy absorbance of the human ear

Peihan Liua,Hang Lib,Wen Jiangc,d,eWen Liuc,d( )Yuehua Qiaoc,d,eHouguang Liua
School of Mechatronic Engineering, China University of Mining and Technology, Xuzhou, 221116, China
Department of Otolaryngology, The Third People's Hospital of Dalian, Dalian, 116033, China
The First Clinical Medical School, Xuzhou Medical University, Xuzhou, 221000, China
Department of Otolaryngology, Affiliated Hospital of Xuzhou Medical University, Xuzhou, 221006, China
Artificial Ear Laboratory of Jiangsu Province, Xuzhou, 221000, China

These authors equally contributed to this work.

Show Author Information

Abstract

This study aims to investigate the impact of middle ear effusion (MEE) on sound transmission in the human ear and its potential diagnostic significance. Firstly, the material properties of specific structures were adjusted based on the existing human ear finite element (FE) model, and the accuracy of the model was validated using experimental data. Secondly, six FE models were developed to simulate varying degrees of MEE by systematically altering the material properties of the middle ear cavity (MEC) at different anatomical locations. Finally, the effects of these six FE models, representing varying degrees of MEE, on sound transmission characteristics and energy absorption (EA) rate in the human ear were systematically analyzed. When the degree of MEE is less than 50% of the MEC volume, its impact on the sound transmission characteristics of the human ear remains minimal, resulting in an estimated hearing loss of approximately 3 dB, with EA rate remaining close to normal levels. Once the effusion exceeds 50% of the MEC volume, a significant deterioration in acoustic transmission is observed, accompanied by a flattening of the EA curve and a drop in EA rates to below 20%. When the effusion completely fills the MEC, the maximum hearing loss reaches 46.47 dB, and the EA rate approaches zero across the entire frequency range. These findings provide theoretical insights into the biomechanical effects of MEE on human auditory transmission and offer a reference for clinical diagnosis and evaluation.

References

【1】
【1】
 
 
Journal of Otology
Pages 176-184

{{item.num}}

Comments on this article

Go to comment

< Back to all reports

Review Status: {{reviewData.commendedNum}} Commended , {{reviewData.revisionRequiredNum}} Revision Required , {{reviewData.notCommendedNum}} Not Commended Under Peer Review

Review Comment

Close
Close
Cite this article:
Liu P, Li H, Jiang W, et al. Numerical analysis of the effect of middle-ear effusion on the sound transmission and energy absorbance of the human ear. Journal of Otology, 2025, 20(3): 176-184. https://doi.org/10.26599/JOTO.2025.9540027

2466

Views

166

Downloads

0

Crossref

0

Web of Science

0

Scopus

0

CSCD

Received: 05 December 2024
Revised: 15 April 2025
Accepted: 16 April 2025
Published: 11 July 2025
© 2025 PLA General Hospital Department of Otolaryngology Head and Neck Surgery. Publishing services by Tsinghua University Press.

This is an open access article under the CC BY-NC-ND license (http://creativecommons.org/licenses/by-nc-nd/4.0/).