Original Research | Open Access

Semiautomated Portable Hand-held 3D Ultrasound Imaging System for Carotid Atherosclerotic Plaque Scanning

Yunqian Huanga,Y, Junni Shia,Y, Chuanjian Chena,Y, Jiawen Lib, Wei Yana, Rui Zhengb( ), Man Chena( )
Department of Ultrasound Medicine, Tongren Hospital, Shanghai Jiao Tong University School of Medicine, Shanghai, China
School of Information Science and Technology, ShanghaiTech University, Shanghai, China

Yunqian Huang, Junni Shi and Chuanjian Chen are contributed equally to this work.

Show Author Information

Abstract

Objective

To develop a portable handhold 3D Ultrasound imaging and diagnostic system, and to validate its reliability and accuracy in the clinical diagnosis of carotid atherosclerotic plaque.

Materials and methods

First acquisition enrolled 55 carotid arteries images by a portable probe which was an important part of the 3D Ultrasound imaging and diagnostic system. 25 arteries were the training set and 30 for validation set. Second acquisition enrolled 84 arteries as an independent validation set. The diagnostic result of our System in carotid plaque including AUC, specificity, sensitivity, accuracy was calculated.

Results

A set of learning-based method for automatic segmentation, reconstruction and diagnosis was established. In the validation set and independent validation set, the AUC, sensitivity, specificity and accuracy achieved 0.708 (95%CI: 0.509-0.908)/0.860 (95%CI: 0.773-0.953), 83.3%/83.9%, 58.3%/88.7%, 73.3%/86.9%, respectively.

Conclusion

The proposed portable handhold 3D Ultrasound imaging and diagnostic system achieved a satisfactory sensitivity, specificity and accuracy in the diagnosis of carotid plaque. Although dependence on operators was an inevitable problem in routine ultrasound examination, the diagnosis results of our system were not highly dependent on examiner experience. In the future, the system could be applied to carotid atherosclerosis screening in communities or remote settings to help address the shortage of ultrasound examiners.

References

【1】
【1】
 
 
Advanced Ultrasound in Diagnosis and Therapy
Pages 193-202

{{item.num}}

Comments on this article

Go to comment

< Back to all reports

Review Status: {{reviewData.commendedNum}} Commended , {{reviewData.revisionRequiredNum}} Revision Required , {{reviewData.notCommendedNum}} Not Commended Under Peer Review

Review Comment

Close
Close
Cite this article:
Huang Y, Shi J, Chen C, et al. Semiautomated Portable Hand-held 3D Ultrasound Imaging System for Carotid Atherosclerotic Plaque Scanning. Advanced Ultrasound in Diagnosis and Therapy, 2026, 10(3): 193-202. https://doi.org/10.26599/AUDT.2026.250087

15

Views

3

Downloads

0

Crossref

0

Scopus

Received: 10 August 2025
Revised: 09 November 2025
Accepted: 04 January 2026
Published: 12 October 2026
2576-2508/© AUDT 2026

This is an open-access article distributed under the terms of the Creative Commons Attribution 4.0 International license, which permits unrestricted use, distribution and reproduction in any medium provided that the original work is properly attributed