Due to the significant correlation between neighboring slices in two-dimensional (2D) multi-slice magnetic resonance data, higher quality slice pictures can be reconstructed by taking use of the redundancy between slices. However, 2D multi-slice magnetic resonance imaging requires an amount of time. To improve the reconstruction quality and speed of 2D multi-slice (MRI) images, proposes a fast 2D multi-slice MRI reconstruction (FMS-JTLHTC) algorithm, which introduces the joint transform learning regular term into the multi-slice hankel tensor completion (MS-HTC) model. Prior to introducing the fast iterative shrinkage-thresholding procedure to accelerate convergence and utilize the graphics processing unit to speed up the procedure, the alternating direction method of multipliers is used to solve the objective issue. Experiments using four brain datasets in two different sampling modes show that the peak signal-to-noise ratio (PSNR) of the FMS-JTLHTC algorithm is improved by an average of 4.04 dB, 3.67 dB, and 2.07 dB compared to the simultaneous atuo-calibrating and K-space estimation (SAKE), low-rank modeling of local K-space neighborhoods with parallel imaging data (PLORAKS) and MS-HTC algorithms, respectively, the reconstruction speed is improved by a factor of 14 compared to the MS-HTC algorithm.