@article{Yang2026, 
author = {Zijian Yang and Yuchen Dong and Weinan Gao and Quan Na and Tengfei Liu},
title = {Output-Feedback Optimal Control to Prevent Hypotension Associated with Labor Epidural Anesthesia},
year = {2026},
journal = {Unmanned Systems},
volume = {14},
number = {3},
pages = {657-667},
keywords = {Labor epidural anesthesia, mean arterial pressure regulation, hypotension, state reconstruction, optimal output-feedback control},
url = {https://www.sciopen.com/article/10.1142/S2301385026410049},
doi = {10.1142/S2301385026410049},
abstract = {In obstetric anesthesia, maintaining the stability of maternal hemodynamic systems is crucial for the safety of both maternal and fetal well-beings. To achieve this goal, manual infusion method is usually used that infuses phenylephrine to patients during surgery. However, traditional manual infusion method may lead to insufficient precision in the dosage. Although automatic control techniques may improve the accuracy of infusion, the implementation of closed-loop pressure control systems usually depends on the full measurement of states, which is hard or impossible to acquire. In this paper, based on auto-regressive with exogenous input models, an output-feedback optimal controller is developed that integrates a state reconstruction technique and an optimal state-feedback control policy. The proposed method effectively compensates for hypotension induced by bupivacaine through precise infusion of phenylephrine. Simulation results demonstrate that the proposed control method can efficiently regulate mean arterial pressure of patients.}
}