Tip optofluidic immunoassay: Evaluating COVID-19 antibody protection with 1 μL fingertip blood

Xiaotian Tan1,2,#( )Yujuan Chai3,#( )Ruihan Li1,2,#Binmao Zhang1,3,#Hao Li1,4,#Jie Zhang5Tianen Zhu6Weishu Wu7Lixiang An4Shi Hu1,2,8Bin Yang5Li Wang5Zhenqiu Cao9Hongjiu Zhang9Peng Wang10Lingling Yu10Shan Yin10Xingyu Li10Fei Shao10Jianheng Huang1,3Jinze Li1Fan Yang2,11Chao Zhao1,2Jiajia Guo1,2Lin Zeng1,2Dong Liang1,2Zhengting Zou9Hairong Zheng1,2( )Xudong Fan7( )Liangzhi Xie5,12,13( )Yunlong Cao10,14( )Hui Yang1,2 ( )
Institute of Biomedical and Health Engineering, Shenzhen Institutes of Advanced Technology, Chinese Academy of Sciences, Guangdong, China
State Key Laboratory of Biomedical Imaging Science and System, Guangdong, China
Department of Biomedical Engineering, Shenzhen University Medical School, Shenzhen University, Guangdong, China
National Innovation Center for Advanced Medical Devices, Guangdong, China
Beijing Key Laboratory of Monoclonal Antibody Research and Development, Sino Biological Inc., Beijing, China
Emergency Department, Shenzhen University General Hospital, Shenzhen University, Guangdong, China
Department of Biomedical Engineering, University of Michigan, Michigan, USA
University of Chinese Academy of Sciences, Beijing, China
Institute of Zoology, Chinese Academy of Sciences, Beijing, China
Changping Laboratory, Beijing, China
Institute of Biomedicine and Biotechnology, Shenzhen Institutes of Advanced Technology, Chinese Academy of Sciences, Guangdong, China
Beijing Engineering Research Center of Protein and Antibody, Sinocelltech Ltd., Beijing, China
Cell Culture Engineering Center, Chinese Academy of Medical Sciences & Peking Union Medical College, Beijing, China
BIOPIC, Peking University, Beijing, China

#These authors contributed equally to this work

Show Author Information

Highlights

• Multimodal COVID-19 antibody protection evaluation with tip optofluidic immunoassay: binding, kinetics, and neutralization.

• Requires only 1 μL of fingertip blood, delivers full results in 40 min; rapid and patient-friendly.

• Lower limit of detection of 0.1 ng/mL, 3–4 log dynamic range, signal-to-noise ratio ~10,000; ensures high precision and sensitivity.

• Identified 4 volunteers with broad-spectrum antibody protection against two strains of SARS-CoV-2.

Abstract

Infectious diseases such as coronavirus disease 2019 (COVID-19) continue to pose significant global health challenges. Effective management of reinfection risks depends on sustained levels of binding and neutralizing antibodies. However, conventional methods—such as enzyme-linked immunosorbent assays (ELISA) and virus neutralization tests (VNT)—are limited by complex workflows, long assay durations, and high sample volume requirements, making them less suitable for routine, decentralized, or time-sensitive surveillance. This study presents a custom-developed tip optofluidic immunoassay (TOI) platform that enables rapid, multiplexed antibody profiling using only 1 μL of fingertip blood. The system integrates batch-fabricated microfluidic immunoreactors with a portable chemiluminescent imaging station, completing both binding and neutralization capability assessments within 40 min. TOI achieves a broad dynamic range (3–4 orders of magnitude), high signal-to-noise ratio (~10,000), and excellent sensitivity for immunoglobulin G (IgG) detection. A renovated version of the rapid in vitro inhibition assay (RIVIA) is incorporated to evaluate neutralizing antibodies against severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) with greater speed and cost-efficiency. In clinical studies, TOI successfully quantified antibody protection against multiple variants, identifying individuals with broad-spectrum immunity to both wild-type and XBB strains. With its high-precision, rapid turnaround, and minimal sample requirement, TOI offers a valuable tool for decentralized immune surveillance and personalized immunization strategy development.

Graphical Abstract

References

【1】
【1】
 
 
hLife
Pages 338-356

{{item.num}}

Comments on this article

Go to comment

< Back to all reports

Review Status: {{reviewData.commendedNum}} Commended , {{reviewData.revisionRequiredNum}} Revision Required , {{reviewData.notCommendedNum}} Not Commended Under Peer Review

Review Comment

Close
Close
Cite this article:
Tan X, Chai Y, Li R, et al. Tip optofluidic immunoassay: Evaluating COVID-19 antibody protection with 1 μL fingertip blood. hLife, 2025, 3(7): 338-356. https://doi.org/10.1016/j.hlife.2025.04.005

642

Views

2

Crossref

Received: 07 February 2025
Revised: 10 April 2025
Accepted: 13 April 2025
Published: 01 July 2025
© 2025 The Author(s).

This is an open access article under the CC BY-NC-ND license (http://creativecommons.org/licenses/by-nc-nd/4.0/).